Operating Room · Fluid Management

Stryker Neptune Waste Management

Closed-system surgical fluid and smoke evacuation unit.

Skantra Resource

AI overview

The Neptune system collects surgical fluids, suctions smoke from electrosurgery, and docks to a hands-free disposal station to eliminate staff exposure to biohazards.

Ask Skantra about this

Common uses

  • Intraoperative suction of blood and irrigation
  • Surgical smoke evacuation
  • Closed disposal at Docking Station

Safety notes

  • Treat all fluid as biohazardous — wear full PPE
  • Only dock at compatible Neptune Docking Station

Setup instructions

  1. 1
    Power on rover; confirm battery and vacuum self-test pass
  2. 2
    Attach a new disposable manifold to the active canister
  3. 3
    Connect suction tubing and smoke evacuation hose
  4. 4
    Verify vacuum levels meet surgical requirement

Troubleshooting

  • Low or no suction

    Check tubing kinks, manifold seal, and canister fill level.

  • Manifold won't seat

    Inspect docking surface; remove debris and reseat firmly.

  • Battery warning

    Return to docking station between cases to recharge.

Training video

My notes

Sign in to save and share notes.